Filmed Grids for TEM
Some tips from our grid coating experts!
The control of the thickness of the coating is extremely important, and it is probably the single most
difficult aspect of the making of coated grids. A "must" in terms of film thickness uniformity is for
the casting liquid surface to be absolutely flat. One should not be picking up film that has formed
too close to the "edge" where the liquid could be concave or convex. If the non-uniformities are
sporadic, then there could be a contamination problem with the liquid.
The procedures will vary with the exact resin used.
For Formvar™ coated grids, 0.25 to 0.3 % Formvar resin
in dry ethylene dichloride. What is not generally appreciated is that there are a number of different grades
of the resin called Formvar, but so far as we know, only one of them is optimum for the filmed grid application.
If one was using other grades, they might require different concentrations or as many have learned, most of the
other grades don't really work at all!
For thicker films, we would suggest increasing the concentration to the range of 0.4 to 0.5%. Thicker films are
usually made when one wants to make a holey film with bigger holes. We find that a concentration above about 0.7%
results in a film that is literally too thick for any EM application.
If one wants thinner films, then reduce the concentration. But if you are already making relatively decent films,
then you are at the borderline in minimum thickness anyhow but the concentration could be reduced a bit more to
about 0.2%. However, below 0.2%, the film becomes just too susceptible to cracking, so that becomes the lower
limit for reducing the thickness and therefore also, the concentration.
We do realize that this is more easily said than done so we provide some additional guidance. Let us assume we
are going to be coating 200 mesh grids. Try to aim for the film, when floating on the water surface, to be
silver to barely silver/gold, in other words, nearly colorless. The support film made from that kind of floating
film should be quite robust in both mechanically, strung over a grid hole, including a slot grid, and also reasonably
stable in the electron beam. In fact the film should be sufficiently robust that a carbon coating should not be
necessary, and this is a real advantage when one is looking at nanoparticles, for example which could be confused
with grain from a carbon film.
But anyone who has made filmed grids will attest, the real test, at the "end of the day", is how the Formvar film
behaves in the vacuum of the TEM and under exposure to the electron beam. These recommended concentrations, and
other aspects of the technique are all optimized in order to produce optimum performance, e.g. long lasting films
with virtually no drift once in the column and being exposed. But one really does need a TEM right there, next
to where the films are being cast, so that they can be instantly inspected and corrective action taken, right
there on the spot, e.g. by way of a concentration change. If this is not done, the the QC step ends up being
when you gear up to do your experiment and then you might very well find out the grids are not usable. Variables
such as temperature, humidity and perhaps other factors seem to come into play, so that the optimum concentration
one day might not quite be the optimum concentration on some other day. There is even one school of thought that
believes that light slowly degrades the solid Formvar which typically might be quite old, since it is used so
sparingly.
The method described above, the one that we believe is used by most makers of filmed grids, is actually only one of
two methods. The other approach involves dipping a glass slide into the Formvar (or other polymer) solution, with
drawing the slide and capturing a thin film of the polymer on the slide. This has also been done with
freshly cleaved mica.
While this approach might sound a bit easier, it too has its difficult "step", that one being the stripping of the
Formvar (or other polymer) film from the slide. In order to facilitate the removal, or stripping of the films, we
would suggest the following procedures:
a. Start with a normal 1" x 3" (25x75 mm) glass microscope slide.
b. Dip the slide into a beaker of saturated Alconox
® solution, which has been previously filtered with Whatman #1 filter paper.
c. Wipe the slide dry with a Kimwipe®
and then, when dry, with several fresh Kimwipes, dry polish the slide by vigorous rubbing. Some experienced film
makers do their rubbing with SPI Lens Tissues or Ross Lens Tissues.
When finished, do a final cleaning with an SPI Two-In-One™ OK Easy
Duster™. Some times veteran coaters claim they have had problems with a particular brand of slides, but
from what we can determine, all brands seem to produce equivalent results, but what can vary is if the particular
slides being used are old, and are exhibiting signs of "glass corrosion". If one experiences problems getting the
films to release, we would suggest getting a fresh box, making sure that it is not from the same batch, but one that
is much more recently manufactured.
Alternative method:
Some veteran experts take the slide in its "as received" condition and dip it into absolute ethanol instead of the
Alconox solution. The slide is then immediately dried with either SPI Lens Tissue or Ross Lens Tissue. One should
immediately proceed with the next step.
d. Apply the Formvar (or other polymer) film, from an ethylene dichloride solution. The Formvar concentration
is a bit higher than mentioned above, and will also depend on the speed with which the slide is withdrawn from the
solution. In this case, the faster the slide is withdrawn, and the faster the liquid is allowed to run off, the
thicker will be the resulting film. This is the part of this technique that causes the most problems because of
the difficulty some persons have in reproducing the conditions of polymer film formation exactly e.g., getting films
of uniform thickness, slide to slide.
e. The polymer film, now on the glass slide should be allowed to dry completely.
f. Holding the slide with the polished slide down, scrape the edges of the slide closest to the polished surface
with a freshly cleaned razor blade. An acetone/alcohol cleaning combination is the recommended approach.
g. Score the polished surface, that is, the one with the polymer film, with the razor blade, and then blow with
an SPI duster any debris that might be remaining.
h. Now using a deep petri dish, one at least several inches deep, and filled nearly to the surface with distilled
water, hold the slide at a 45° angle to the water surface, and then slowly lower it into the water itself. When the
water reaches the score marks (parallel to the short dimension of the slide), very slowly continue to lower the slide
and surface tension effects should then do the rest, with the polymer film smoothly floating off of the slide.
i. At that point, grids can be picked up, one at a time, and dunked into the water and then brought up underneath
the floating film and with lifted out of the water and dried.
This second procedure should also result in excellent Formvar or other filmed grids. What makes this part of the
process so interesting is that a technique one person finds easy to practice, another finds difficult and
vice versa. If you are still encountering difficulties with the Formvar removal from the slide,
there are other alternatives.
We would caution the unitiated, however, that despite the ease at which it sounds one can make filmed grids of
high quality, just remember that this is indeed an art and one does not learn an art in a few minutes. Be prepared
to spend some number of hours perfecting your own art in the making of filmed grids for TEM! We have people in
our own laboratory who have been making support films for as much as 50 years, and even they have to adjust the
dilution, etc. to get films that are consistently beam stable.
To Ask a Question or Make a Comment
To Place an Order or Request a Quote
Return to:
© Copyright 1998 - . By Structure Probe, Inc.
Contacting SPI Supplies and Structure Probe, Inc.
All rights reserved.
All trademarks and trade names are the property of their respective owners.
Worldwide Distributors, Representatives, and Agents