SIRA Calibration
Specimen Set

Complete Use Instructions



General Information about the SIRA Specimen

These high quality calibration specimens provide a means of testing scanning electron microscopes, scanning transmission electron microscopes (in the secondary electron mode), and electron probe microanalysis systems particularly for magnification accuracy above 20X, for tilt specifications on the stage, for image distortion and for the measurement of the depth of field. However they are also usable for tests of stage stability, centering of stub rotation, checks for symptoms of electrical and mechanical interference, etc.

Two metal (resin-backed) grating replicas, one of a 19.7 and the other of a 2160 lines/mm original are mounted on sample mounts and gold coated. Each replica contains lines in two orthogonal directions with line frequencies guaranteed to be within 1% of the stated figure on delivery. The specimens are flat with a usable area of over 60 square mm, and were originally developed by SIRA EM Techniques Group.

For more detailed information, the reader is referred to an early research note: "A New Magnification Test Specimen for SEMs" by I. M. Watt and N. A. Wright, Metron (UK) 3, No 6, 153-156 (1971).

Using the Specimen

1. Magnification Calibration


The 19.7 lines/mm sample (Figure 1) is suitable from about 20X to 1600X magnification on the instrument "CRT"; the finer specimen ruled to 2160 lines/mm (Figure 2) can be used from 1000X and up. It is essential that the user first ensures that the image distortion levels are within acceptable tolerances at the magnification levels of interest before embarking on the magnification calibration. In addition, on some instruments it is necessary to work the magnification calibration routine in well defined and reproducible operational conditions (e.g. working distances, KV). Further it is better to work with zero tilt in the direction used for calibration. For example, Figures 1 and 2 were recorded with X-tilt on the stage but zero Y tilt. The calibration was performed using only the vertical lines which are not disturbed by the X-tilt. Magnification calibration routines should include checks on recorded images since the recording video display unit (CRT) may give a different magnification than the visual CRT.


The 19.7 lines/mm sample, use from about 20-1600X

The 2160 lines/mm sample, use from about 1000X and upwards


2. Calibration of the stage tilt controls

At normal incidence of the electron probe on the specimen (zero X- and Y- tilt), the unit cells in the pattern are square. They become rectangular as the specimen plane is inclined to the electron-optical axis.

Figure 3 shows this effect for the case of the X-tilt, that is, tilt around the Y-axis. The ratio of short to long dimensions of the rectangle is the cosine of the actual tilt angle provided that the tilt axis is parallel to one set of the ruling. The coarse (19.71 lines/mm) sample is recommended for tilt-angle calibration.

Demonstration of the effect
of X-tilt around the Y- axis

3. Observation of image distortion in the microscope

The microscopist might want to measure and/or correct the degree of image distortion at various working distances, magnifications, and operational KV levels, normal incidence Figure 4 is recommended. It is also important to remember that distortions arise all along the imaging chain to the recording CRT. Hence it should not be forgotten that recorded images are important at this point and should never be over looked.

Measure and correct the degree of distortion
at different working distances, magnifications,
and KV settings.

4. Measurement of depth of field

Use a moderate (30 - 40o) X-tilt to demonstrate or check the depth of field by assessing the Z-depth for good focus. The depth of field is higher when the final aperture is smaller, when the working distance is longer, and when the SEM is operated in the lower magnification ranges.

To Ask a Question or Make a Comment

To Place an Order or Request a Quote


Return to:
  • Standards and Calibration Aids
  • SPI Supplies Catalog Table of Contents
  • SPI Supplies Home Page.

  • Monday March 22, 2010
    © Copyright 1998 - 2010. By Structure Probe, Inc.
    Contacting SPI Supplies and Structure Probe, Inc.
    All rights reserved.
    All trademarks and trade names are the property of their respective owners.
    Privacy Policy

    Worldwide Distributors, Representatives, and Agents Flag logo