SPI Supplies is the source!

Eyepiece Reticles from Graticules Ltd.

Asbestos fiber analysis




SPI Supplies offers the full range of eyepiece reticles from Graticules Ltd. designed for asbestos fiber analysis.

In general, the total magnification of your light microscope can be determined by simply multiplying the objective power by that of the eyepiece.

For example:
Objective Eyepiece Total Magnification
20x 10x 200x


For most applications, especially with modern optics, this is sufficient. However, where actual size is critical, it is often times necessary to modify the dimensions, so when ordering an eye piece reticle, the following information is needed:

To calibrate the instrument, fit the microscope with an eyepiece scale and appropriate stage micrometer by Graticules Ltd.

Compare the length of the stage micrometer with the eyepiece scale. An exact calibration factor can be determined.

Example Calculation:
Using a 40x objective lens, a circle in the eyepiece requires a diameter of 4000 µm (4 mm) to coincide with or read a 100 µm circle on the stage. The factor is therefore defined as 4.


To Ask a Question or Make a Comment

To Place an Order or Request a Quote


Return to :