
SPI Critical Point Drying Apparatus
Detailed Description of the Apparatus and Service Information:
A Complete Run
|
- Wash tissue (see previous section for details).
- Fix tissue.
- Dehydrate with acetone or ethanol.
- If using ethanol, substitute with amyl acetate or Freon® 113.
- Place tissue in baskets taking care not to allow drying;
this operation should be carried out under the solvent.
- Fill transfer boat with transfer liquid (acetone, amyl
acetate, Freon® 113) and put baskets in boat.
- Ensure apparatus is correctly installed (Section 2.1.3) and
run cold water to cool chamber to 20° C.
- Load specimen boat, close door and open inlet valve.
Filling with liquid gas should be rapid; open vent valve to
avoid back-pressure.
- Leave inlet valve fully open (with vent valve slightly open
to maintain liquid level) and open drain valve to remove
substitution liquid. This flushing action should be kept up
for 3-5 minutes. If the noise is found to be objectionable,
a 'silencer' in the form of a length of plastic pipe may be
fitted to the drain valve; this also serves to direct the
flow of exhausting solvent.
- After flushing the bulk of the substitution liquid, fill the
chamber and close all valves. Leave for 1 hour to allow
impregnation. (The times mentioned refer to a piece or
pieces of tissue less than 1mm in thickness. For larger
pieces or very dense tissue allow longer time).
- Repeat operation 9.
- Close the inlet valve and allow the level of liquid to fall
to about the level of the top of the boat.
- Complete the drying run (see page 13 for details) by heating
slowly to 36-38° C WITH ALL VALVES CLOSED.
- Carefully vent the carbon dioxide gas. Avoid condensation
effects. NOTE: Condensation can be caused by rapid
venting or by the presence of acetone or amyl acetate.
- Remove the tissue baskets from the chamber, mount the
specimens on microscope stubs and proceed with vacuum
coating.
To Ask a Question or Make a Comment
To Place an Order or Request a Quote
Return to:
Friday August 29, 2008
© Copyright 2000 - 2008. By Structure Probe, Inc.
Contacting SPI Supplies and Structure Probe, Inc.
All rights reserved.
All trademarks and trade names are the property of their respective owners.
Privacy Policy
Worldwide Distributors, Representatives, and Agents