All of the trouble shooting indicators listed above in section 6 apply to membrane incubations. In addition the following should be considered for membrane incubations.

(I) Poor labeling
(A) Check transfer system is working.

(B) High background is masking specific staining.


(II) Excessive background
(A) Inadequate blocking of transfer membrane (especially nylon based membranes). Saturate for longer times at higher temperatures. Use higher concentrations of BSA in the gold conjugate. Include 0.1% gelatin in the gold conjugate to reduce binding to sulphur. Wash the membrane very thoroughly before the incubations.

(B) Spontaneous precipitation of silver due to impure water used for washing membrane. Use only clean distilled water in clean trays with no contamination. Use a totally negative control simultaneously. Inadequate washing of membrane following the gold incubation. Lengthen the washing procedure in water to remove all salts.